Renew of registration of domestic Class III medical devices
2013-12-05

Ⅰ.Project name: Domestic Medical Device Registration

Ⅱ.Contents of licensing: renew of registration of domestic Class III medical devices

Ⅲ.Basis for implementation: "Regulations for Supervision and Administration of Medical Devices", "Provisions for Medical Device Registration"

Ⅳ.Fees: No charge

Ⅴ.Quantity restrictions: No quantity limit for the licensing

Ⅵ.List of application documents:
Data number (1) Initial registration form for domestic Class III medical devices;
Data number (2) Medical device manufacturer certification;
Data number (3) The original registration certificates of medical devices;
Data number (4) product registration test report issued by medical device testing institutions;
Data number (5) Applicable product standards and instructions;
Data number (6) Product quality tracking reports;
Data number (7) Medical device manuals;
Data number (8) valid certifying documents for evaluation (certification) of production quality system - quality system evaluation report for specific product should be provided accordingly;
Data number (9) For circumstances described in Article 34 of the "Provisions for Medical Device Registration", the applicant shall provide the corresponding explanations and certifying documents;
Data number (10) Self-assurance statement of authenticity of the dossiers submitted;
Data number (11) Other issues.

VII.Requirements for application dossiers:
(I)General requirements for application dossiers:
The first page of the application dossiers shall be a directory of the application items, the order of which shall be arranged as per by Annex v of the "Provisions for Medical Device Registration". Each dossier shall have a cover page indicating the name of the product and the applicant, and the dossier title shall be marked at the upper right corner. The dossiers shall be distinguishable with clear markers, and the name and serial number of the dossier shall be indicated. The entire dossiers shall be bound into a book.
The application dossiers shall be in One set and printed on A4 paper, the content shall be complete, clear, and shall not be altered, the documents issued by the Government and other institutions shall be provided in the original size.
Copies of the dossiers should be clear.
The product names in various items (approvals for marketing, standards, test reports, manuals) of applications dossiers shall correspond with the names and substantial contents in the Application form. Trade name (if any) should be marked. The application dossiers shall be in Chinese, and translated dossiers shall be accompanied with the originals.

(II)Specific requirements for application dossiers:
1.Initial registration form for domestic Class III medical devices;
Medical device registration application form and the "2010 Electronic medical devices (IVD) application software (including manuals) "can be downloaded at www.cfda.gov.cn;
Medical device registration application form is one of the important data for registration application, therefore the items shall comply with the requirements of instructions.

2.Medical device manufacturer certification;
Copies of the Production License and Business License, stamped with the seal of the enterprise;
The applied product should be within the scope circumscribed by the production license;
And within the validity period.

3.The original registration certificates of medical devices;
(1)For circumstances described in Article 33, Chapter 5 of the "Provisions for Medical Device Registration", a copy of the original registration certificate of the medical device shall be submitted;
(2) For renew of registration due to changes, the manufacturer shall submit a copy of the original registration certificate of the medical device, which shall be marked “identical with the original” and stamped by the manufacturer; and the original certificate shall be returned while receiving the new ones.

4.Product registration test report issued by medical device testing institutions;
The specifications and models of the tested products should be within the scope of the registration application;
Test type should be registration test or national full-performance sampling test;
Originals;
Should be within the validity period (as per Article 4 in Annex V of the "Provisions for Medical Device Registration").
Note: If the Article 11, 12 and 13 of the "Provisions for Medical Device Registration" are implemented, the manufacturer shall provide appropriate documentation stamped with its official seal.

5.Applicable product standards and instructions;
Standard text shall be stamped with the official seal of the manufacturer;
Development instructions (applicable to registered product standards);
Product standards and development instructions shall be in duplicate, together with two statements of standards consistency (applicable to registered product standards);
The applied products should be included in the product standard range;
If national and industry standards are employed as applicable product standards:
①The manufacturer shall provide a statement of compliance with national and industry standards, which shall be stamped with the official seal of the manufacturer;
②A statement for bearing full responsibilities of the quality of post-marketed products produced by manufacturer, which shall be stamped with the official seal the manufacturer;
③A description of product models and specification classifications produced by manufacturer, which shall be stamped with the official seal the manufacturer.

6.Product quality tracking reports;
Should be stamped with the official seal of the manufacturer.

7.Medical device manuals;
Product manual should be provided and should be stamped with the seal of the manufacturer; the omission of product manual shall be explained by the manufacturer’s specified documents, and stamped with the official seal of the manufacturer.

8.Valid certifying documents for evaluation (certification) of production quality system - quality system evaluation report for specific product should be provided accordingly;
(1)Stamps of the (food) drug administration departments of the Provinces, autonomous regions and municipalities; a copy of medical device quality system certificate (if any) can be submitted, but it should be stamped with the official seal of the holder;
(2)And within the validity period.
(3)Product system shall cover the applied product

9.For circumstances described in Article 34, Chapter 5 of the "Provisions for Medical Device Registration", the applicant shall submit corresponding supporting documents stamped with the official seal of the manufacturer.

10.Self-assurance statement of authenticity of the dossiers submitted;
List of submitted dossiers;
Manufacturers commitment for legal liabilities;
Should be stamped with the official seal of the manufacturer.

11.Others
(1)For renew of registration without any changes occurred in products, product standards and manuals, the manufacturer shall submit a Statement of No Change and no longer need to submit registration product standards, test reports and manuals, which shall be maintained after obtaining approval;
(2)For renew of registration with changes occurred in products, product standards or manuals, the manufacturer shall submit, in addition to the required application dossiers, a comparison table of the products, product standards or manuals before and after the change, as well as the corresponding supporting information.

VIII.Application process diagram:

IX.Approval procedures:
(I)Acceptance:
After the application is filed to the Administrative Acceptance Service Center, and the application dossier submitted is in accordance with the list in VI of this Notice, the format examination shall be conducted by the work staff of the Center according to the requirements of the "Acceptance standards for domestic Class III and overseas medical device registration application dossiers"(SFDA Department of medical device supervision [2005] No. 111). Where the application item is not subject to administrative approval according to law, a decision of non-acceptance shall be notified to the applicant immediately. Where the application item is not within the functions of the administrative authority, a decision of non-acceptance shall be made immediately, and the applicant shall be informed to launch the application to relevant administrative authority. Where in the application dossier, there are errors that can be corrected on the spot, the applicant shall be allowed to make the corrections on the spot. Where the application dossier is incomplete or does not meet the statutory format, the applicant shall be informed of all the supplements and amendments to be made once and for all on the spot or within five days. If the applicant is not informed within the above-mentioned period, the application shall be deemed as accepted from date of receipt of application materials. Where the application dossier is within the functions of the administrative authority, the application materials are complete and comply with the statutory format, or where the applicant has submitted all supplements and amendments in accordance with the requirements of the administrative authority, the application for administrative approval shall be accepted.

(II)Review:
Once accepted, Administrative Service Center shall deliver the application dossiers to the Center for Medical Device Evaluation (CMDE) for technical review, which includes product testing and expert review, and shall not exceed 60days. But the time for the applicant’s rectification, which is based upon opinions after expert review, shall not be included in the time frame for licensing.

(III)Licensing decisions:
Upon receipt of the dossiers after CMDE technical review, CFDA shall make a decision for registration within 30 days, and written explanations shall be made for rejection of registration.

(IV)Delivery:
Within 10 days from the date of decision making of administrative examination and approval, CFDA Administrative Service Center shall deliver the decision to the applicants.

X.Commitment time frame:
The administrative licensing decision shall be made within 90 days from the date of acceptance.

XI.Authority of Implementation:
Implemented by: CFDA
Accepted at: CFDA Administrative Service Center

XII.Changes:
If any content listed in medical device registration certificate changes, the certificate holder shall, within 30 days from the date of change, apply for changes procedures or renew of registration.

XIII.License validity and renewal:
Medical device registration certificate is valid for four years. Upon expiry of Medical Device Registration Certificate, if the sales or use of medical devices are still necessary, the manufacturer shall apply for renew of registration within six months before the valid date. If the manufacturer failed to do so, registration testing shall be required for renew of registration.

XIV.Annual inspection or annual review of the license: None

XV.Institutions for inquiries and complaints:
Inquiries: CFDA
Complaints: CFDA Bureau of Investigation and Enforcement, Department of Legal Affairs
Note: The time frame of this TIPS counts on working days, excluding legal holidays